Home » Products » Load Cells » Model i100
Email

Model i100

Strain Gauge Transducer Simulator

i100.jpg
Download product tech sheet
To view in your browser: Left-Click on the icon. To save: Right-Click on icon / Save Target As
The i100 hand held simulator generates precise mV/V signals to help develop, troubleshoot, and
calibrate strain gage signal conditioners, instruments, signal processors, and data loggers.

Accuracy is 0.03%, temperature effects are 6 ppm/F, and zero balance is 0.0004 mV/V. Resistance of the i100 simulator is equivalent to a 350-ohm bridge. The i100 has 8 switch selectable output steps.

Also, a vernier knob is provided to allow the user to continuously adjust the output from 2 to + 2 mV/V. A convenient switch provides true reverse polarity. Connection to the i100 is made through either a PT style connector or 4 color-coded spring-loaded test clips. All critical internal contacts are gold plated. The i100 is compatible with AC carrier or DC strain gage signal conditioner electronics.
Output (mV/V)
0. 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5 & variable
Input (V)
10 optimum, 15 max
Accuracy (%)
0.03
Zero Balance (mV/V)
0.0004 @ 22 deg C and zero output
Temperature Coefficient
10 ppm/deg C
Resistance (ohms)
350